SUPPLEMENT TO
INSTALLATION &
STARTUP MANUAL:
PROCEDURE FOR
ELEVATED
TEMPERATURE SENSOR
HEAD INSTALLATION
CiDRA Corporate Services 50 Barnes Park North Wallingford, CT 06492
Tel. 203-265-0035 Fax. 203-294-4211 www.cidra.com
Vue de la page 0
1 2 3 4 5 6 ... 21 22

Résumé du contenu

Page 1

SUPPLEMENT TO INSTALLATION & STARTUP MANUAL: PROCEDURE FOR ELEVATED TEMPERATURE SENSOR HEAD INSTALLATION CiDRA Corporate Services 50 Barne

Page 2

Remove the springs from the alignment sheet Bundle the individual sensor wires and wires #9 and #10 together and attach them to the alignment sheet us

Page 3 - List of Figures

2.4 Sensor Cover Installation 2.4.1 Cover Seal Installation Note: The seals to be installed on each cover half must seal on each other when the cover

Page 4

Repeat the seal installation on the opposite cover half. Note: Ensure the seals are installed to mirror each other as they must seal on each other d

Page 5 - 1 INTRODUCTION

2.4.2 Installing Cover on Pipe Remove the pre-amplifier housing cover from the upper cover assembly. Disconnect external connector cable assembly fr

Page 6 - 2 SENSOR HEAD INSTALLATION

For horizontal pipe installations, install the cover half with the preamplifier box over the sensor band such that it is centered on the band and over

Page 7 - CAUTION

115 11735913 191720 1862 4814 10 12 16 Install 5/16”–18 cover bolts fitted with a washer and lock washer in each cover bolt hole numbered 1 – 16 i

Page 8

2.4.3 Cover End Seal Installation Un-spool a length of adhesive backed seal and wrap it around the pipe. Overlap the length of seal by about 1.5 inc

Page 9 - 2.3.3 Sensor Installation

Repeat this installation procedure on the opposite end of the cover. Install the cover end seal compression rings so that they overlap the cover flang

Page 10

2.5 Sensor Pre-Amplifier Connections 2.5.1 Sensor Band to Pre-Amplifier Connection Carefully feed the eight sensor wires and wires #9 and #10 from t

Page 11

WARNING Discharge each sensor per user manual instructions just prior to connecting sensors to preamp. Failure to do so may result in damage to prea

Page 13

2.5.2 Sensor to Transmitter Cable Connection Install the sensor cover to transmitter cable connector. Terminate the individual sensor to transmitter

Page 15 - 21020-01 Rev 01 Page 2-11

CiDRA Corporate Services 50 Barnes Park North Wallingford, CT 06492 (In U.S.): 877-cidra77 Tel: 203-265-0035 Fax: 203-294-4211 Visi

Page 16

Table of Contents 1 INTRODUCTION...1-1 2

Page 18

1 INTRODUCTION This manual is intended to be a supplement to the SONARtrac® Process Monitoring System Installation & Startup Manuals. The inform

Page 19 - WARNING

2 SENSOR HEAD INSTALLATION 2.1 Caution: High Temperatures Be careful when installing or removing the high temperature system. Incidental contact wi

Page 20 - 21020-01 Rev 01 Page 2-16

2.3 Sensor Installation 2.3.1 Sensor Insulator Installation Properly clean the process pipe as described in the VF-100 or GVF-100 Manual along a 36 i

Page 21

Note: Wires #9 and #10 on the alignment sheet are used for grounding and temperature. XXXXXXXX Alignment sheet Alignment sheet retaining springs Kap

Page 22 - CiDRA Corporate Services

2.3.3 Sensor Installation Slightly bend the alignment tabs on the sensor band alignment sheet upward. These will serve to align the sensors during in